An optical fiber arterial pulse wave sensor is proposed using an in-line Michelson interferometer that is a hollow optical fiber spliced to a single-mode fiber at one end and cleaved at the other end. The proposed optical fiber arterial pulse wave sensor consists of an in-line Michelson interferometer and steel reinforcement enclosed in a heat-shrinkable tube. The sensor was directly attached onto a wrist and signals corresponding to arterial pulse waves successfully obtained. The signal-to-noise ratio of the sensor signals was better than 20 dB.
| Original language | English |
|---|---|
| Pages (from-to) | 1318-1321 |
| Number of pages | 4 |
| Journal | Microwave and Optical Technology Letters |
| Volume | 52 |
| Issue number | 6 |
| DOIs | |
| State | Published - Jun 2010 |
Keywords
- Arterial pulse waves sensor
- In-line Michelson interferometer
- Optical fiber sensor