Abstract
We present a SMART injector with two parallel common-path optical coherence tomography fibers to enable angle measurements and injection depth corrections for oblique subretinal injection. The two optical fibers are attached to opposite sides of a 33 G needle with known offsets and designed to pass through a 23 G trocar that has an inner diameter of 0.65 mm. By attaching a SMART system to a rotational stage, the measured angles are calibrated for minimal error from reference angles. A commercial eye model was used to evaluate the control performance, and injection experiments were performed on a phantom made of agarose gel and a porcine eye.
| Original language | English |
|---|---|
| Pages (from-to) | 926-939 |
| Number of pages | 14 |
| Journal | Biomedical Optics Express |
| Volume | 12 |
| Issue number | 2 |
| DOIs | |
| State | Published - Feb 2021 |
Bibliographical note
Publisher Copyright:© 2021 Optical Society of America under the terms of the OSA Open Access Publishing Agreement.