Oblique injection depth correction by a two parallel OCT sensor guided handheld SMART injector

Jintaek Im, Cheol Song

Research output: Contribution to journalArticlepeer-review

10 Scopus citations

Abstract

We present a SMART injector with two parallel common-path optical coherence tomography fibers to enable angle measurements and injection depth corrections for oblique subretinal injection. The two optical fibers are attached to opposite sides of a 33 G needle with known offsets and designed to pass through a 23 G trocar that has an inner diameter of 0.65 mm. By attaching a SMART system to a rotational stage, the measured angles are calibrated for minimal error from reference angles. A commercial eye model was used to evaluate the control performance, and injection experiments were performed on a phantom made of agarose gel and a porcine eye.

Original languageEnglish
Pages (from-to)926-939
Number of pages14
JournalBiomedical Optics Express
Volume12
Issue number2
DOIs
StatePublished - Feb 2021

Bibliographical note

Publisher Copyright:
© 2021 Optical Society of America under the terms of the OSA Open Access Publishing Agreement.

Fingerprint

Dive into the research topics of 'Oblique injection depth correction by a two parallel OCT sensor guided handheld SMART injector'. Together they form a unique fingerprint.