Magnetic sensor-based detection of picoliter volumes of magnetic nanoparticle droplets in a microfluidic chip

Ilgyo Jeong, Young Jae Eu, Kun Woo Kim, Xing Hao Hu, Brajalal Sinha, Cheol Gi Kim

Research output: Contribution to journalArticlepeer-review

10 Scopus citations

Abstract

We have designed, fabricated and tested an integrated microfluidic chip with a Planar Hall Effect (PHE) sensor. The sensor was constructed by sequentially sputtering Ta/NiFe/Cu/NiFe/IrMn/Ta onto glass. The microfluidic channel was fabricated with poly(dimethylsiloxane) (PDMS) using soft lithography. Magnetic nanoparticles suspended in hexadecane were used as ferrofluid, of which the saturation magnetisation was 3.4 emu/cc. Droplets of ferrofluid were generated in a T-junction of a microfluidic channel after hydrophilic modification of the PDMS. The size and interval of the droplets were regulated by pressure on the ferrofluid channel inlet. The PHE sensor detected the flowing droplets of ferrofluid, as expected from simulation results. The shape of the signal was dependent on both the distance of the magnetic droplet from the sensor and the droplet length. The sensor was able to detect a magnetic moment of 2× 10-10 emu at a distance of 10 μm. This study provides an enhanced understanding of the magnetic parameters of ferrofluid in a microfluidic channel using a PHE sensor and will be used for a sample inlet module inside of integrated magnetic lab-on-a-chip systems for the analysis of biomolecules.

Original languageEnglish
Pages (from-to)302-307
Number of pages6
JournalJournal of Magnetics
Volume17
Issue number4
DOIs
StatePublished - Dec 2012

Keywords

  • Droplet
  • Ferrofluid
  • Magnetic nanoparticle
  • Magnetic sensor
  • Microfluidics

Fingerprint

Dive into the research topics of 'Magnetic sensor-based detection of picoliter volumes of magnetic nanoparticle droplets in a microfluidic chip'. Together they form a unique fingerprint.

Cite this