Abstract
High frequency ultrasound over 40 MHz has been used to image the anterior segment of the eye, but it is not suitable for the posterior segment due to the frequencydependent attenuation of ultrasound and thus the limitation of penetration depth. This paper proposes a novel scan method to image the posterior segment of the eye with an angled high frequency (beyond 40 MHz) ultrasound needle transducer. In this method, the needle transducer is inserted into the eye through a small incision hole (̃1 mm in diameter) and rotated around the axial direction to form a cone-shaped imaging plane, allowing the spatial information of retinal vessels and diagnosis of their occlusion to be displayed. The feasibility of this novel technique was tested with images of a wire phantom, a polyimide tube, and an excised pig eye obtained by manually rotating a 40-MHz PMN-PT needle transducer with a beveled tip of 45°. From the results, we believe that rotational scan imaging will help expand the minimally invasive applications of high frequency ultrasound to other areas due to the capability of increased closeness of an angled needle transducer to structures of interest buried in other tissues.
Original language | English |
---|---|
Article number | 4816075 |
Pages (from-to) | 676-680 |
Number of pages | 5 |
Journal | IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control |
Volume | 56 |
Issue number | 3 |
DOIs | |
State | Published - Mar 2009 |
Bibliographical note
Funding Information:Manuscript received august 12, 2008; accepted october 11, 2008. This work was supported by Mr. louis Fox with a private grant and the support of nIh grant # P41-EB2182 and of the Korea research Foundation grant # KrF-2008-013-c00036. d.-G. Paeng is with cheju national University, Marine Industrial Engineering, Jejudaehakno 66, Jeju, 690-756, south Korea.