Electromagnetic compatibility of two novel packaging concepts of an inductively powered neural interface

  • Sohee Kim
  • , Martin Wilke
  • , Matthias Klein
  • , Michael Toepper
  • , Florian Solzbacher

Research output: Chapter in Book/Report/Conference proceedingConference contributionpeer-review

6 Scopus citations

Abstract

For implantable neuroprosthetic devices, reliable packaging of the device is one of the requirements. Selection of proper packaging material and design is essential to protect the device from moist fluidic environment in the body as well as not to cause any harm to the implanted tissue. In particular, packaging needs to be designed carefully when the device operates wirelessly through inductive coupling between two coils, since the packaging could degrade the power transmission due to the losses in the packaging itself. In this study, the influence of the proposed packaging methods on the magnetic power transmission was investigated using numerical simulations with the help of finite element analysis. The simulation showed that both proposed packaging concepts of using a silicon lid and LTCC ferrite lid do not degrade the power transmission significantly at the selected frequency of 2.64 MHz.

Original languageEnglish
Title of host publicationProceedings of the 3rd International IEEE EMBS Conference on Neural Engineering
Pages434-437
Number of pages4
DOIs
StatePublished - 2007
Event3rd International IEEE EMBS Conference on Neural Engineering - Kohala Coast, HI, United States
Duration: 2 May 20075 May 2007

Publication series

NameProceedings of the 3rd International IEEE EMBS Conference on Neural Engineering

Conference

Conference3rd International IEEE EMBS Conference on Neural Engineering
Country/TerritoryUnited States
CityKohala Coast, HI
Period2/05/075/05/07

Fingerprint

Dive into the research topics of 'Electromagnetic compatibility of two novel packaging concepts of an inductively powered neural interface'. Together they form a unique fingerprint.

Cite this